User Manual
Operation Multi 3500i
46
Preparatory activities Perform the following preparatory activities when you want to calibrate:
Carrying out an
ISE calibration
Proceed as follows to calibrate the instrument:
1 Connect the ISE combination electrode to the measuring in-
strument. The pH/mV/ISE measuring window is displayed.
2 Keep the standard solutions ready.
3 Measure the temperature of the standard solutions using a
thermometer.
1 In the measured value display, select the ISE measuring win-
dow with <▲><▼> and <M>.
2 Start the calibration with <CAL>.
The calibration display appears.
3 Thoroughly rinse the electrode with distilled water.
4 Immerse the electrode in standard solution 1.
5 Select the Set standard setting with <▲><▼> and press
<MENU/ENTER>.
6 Set the concentration of the standard solution with <▲><▼>
and press <MENU/ENTER>.
7 Measure the temperature of the standard solution using a ther-
mometer.
8 Select the Set temperature setting with <▲><▼> and press
<MENU/ENTER>.
9 Set the temperature with <▲><▼> and press
<MENU/ENTER>.
10 Select Continue with <▲><▼> and press <MENU/ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
i Immerse sensor in std. 1
Set temperature:24 °C
Continue
Set standard: 0.010 mg/l
ISE <CAL>