Multi 3500i What to do if...
83
6 What to do if...
6.1 pH and ORP measurement
Error message,
OFL or UFL
(measuring range
exceeded or undercut)
Error message, Error
(calibration error)
Cause Remedy
Electrode:
Air bubble in front of the
diaphragm
Remove air bubble
Air in the diaphragm Extract air or
moisten diaphragm
Gel electrolyte dried out Replace electrode
Test sample
The pH value lies outside the
measuring range
not possible
Cause Remedy
Electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Moisture in the plug Dry plug
Not enough electrolyte Top up electrolyte
Electrode obsolete Replace electrode
Electrode broken Replace electrode
Socket damp Dry socket
Calibration procedure:
Incorrect solution temperature
(without temperature sensor)
Set up correct temperature
Incorrect buffer solutions Select buffer solutions
suitable for the calibration
procedure
Buffer solutions too old Use only once.
Note the shelf life