User Manual

What to do if... Multi 3500i
84
No stable measured
value
Obviously incorrect
measured values
Sensor symbol flashes
Cause Remedy
pH electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Test sample
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
pH electrode:
Not connected Connect electrode
Cable broken Replace electrode
pH electrode unsuitable Use suitable electrode
Temperature difference
between buffer and test
sample too high
Adjust temperature of buffer
or sample solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Calibration interval expired Recalibrate the measuring
system