Multi 3500i What to do if...
85
6.2 ISE measurement
Error message OFL
Obviously incorrect
measured values
Error message, Error
(invalid
calibration)
Warning [TempErr]
Warning [ISEErr]
Cause Remedy
Measuring range exceeded Dilute test sample
Cause Remedy
Electrode not connected Connect electrode
Cable broken Replace electrode
Cause Remedy
Ion-sensitive electrode:
Moisture in the plug Dry plug
Electrode obsolete Replace electrode
Electrode unsuitable for the
range to be measured
Use a suitable electrode
Socket damp Dry socket
Calibration procedure:
Wrong sequence of standards
for three point calibration
Select correct sequence
Calibration standards do not
have the correct temperature
(max.
± 2 °C temperature
difference)
Adjust the temperature of
the calibration standards
Cause Remedy
Temperature difference
between measurement and
calibration greater than 2 K.
Adjust the temperature of
the test sample
Cause Remedy
Electrode voltage outside
calibrated range
Recalibrate