What to do if... Multi 3500i
86
6.3 DO measurement
DO sensor was not
recognized
Error message OFL
Error message, Error
(invalid calibration)
Error message, LEAK
Measured value flashes
for a long time
(No stable measured
value)
Measured value too low
Cause Remedy
Sensor not connected Connect sensor
Cable broken Replace sensor
Cause Remedy
Measuring range exceeded not possible
Cause Remedy
DO sensor:
Electrolyte solution depleted Regenerate sensor
Membrane contaminated Clean membrane
Electrode system poisoned Regenerate sensor
Cause Remedy
DO sensor:
Membrane damaged Exchange the membrane cap
Membrane head screwed on
not tight enough
Screw membrane head tight
Cause Remedy
DO sensor:
Membrane contaminated Clean membrane
Cause Remedy
DO sensor:
Insufficient flow Provide flow to the sensor