User Manual

Multi 3500i What to do if...
87
Measured value too
high
Sensor symbol flashes
6.4 Conductivity measurement
Conductivity
measuring cell
was not recognized
Error message, Error
(invalid
calibration)
Sensor symbol flashes
Cause Remedy
High amount of dissolved
substances
Correct solubility function
using the salinity equivalent
Air bubbles bump on the
membrane with high velocity
Avoid direct flow to the
membrane
The carbon dioxide pressure is
too high (> 1 bar)
Measuring not possible
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and replace it if
necessary
Unsuitable control standard Use control standard,
0.01 mol/l KCl
Cause Remedy
Calibration interval expired Recalibrate the measuring
system