4.3.2 Calibrating via a comparison measurement
Proceed as follows to calibrate the instrument:
1 Connect the D. O. sensor to the measuring instrument.
2 Press the <CAL> key repeatedly until O
2
FAC appears.
3 Immerse the D. O. sensor in the calibrating solution.
4 Press the <RUN/ENTER> key.
The AutoRead measurement begins. If the measured value is
stable, the instrument displays the determined measured value
and the correction factor of the relative slope (initial value =
1.000).
5 Remove the D. O. sensor from the calibrating solution.
6 Perform a comparison measurement, e. g. Winkler titration,
with the calibrating solution (determine the nominal value).
Leave the measuring instrument switched on until the nominal
value is determined.
7 Set the displayed value on the measuring instrument to the
nominal value with <▲> <▼>. This changes the related correc-
tion factor of the relative slope.