Manual

Operation Oxi 1970i
44
ba75691e01 01/2007
4.7 Reset
You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
Measurement
parameters
The following measured parameters (O
2
InI) are reset to the default
condition:
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery
status:
Measuring mode D. O. concentration
AutoRange automatic measure-
ment range selection
On (YES)
Relative slope 1.00
Correction factor of the relative
slope
1.000
Salinity equivalent 0.0
Salinity correction Off
Baud rate 4800
Interval 1
(automatic saving) OFF
Interval 2
(for data transmission) OFF