Manual

What to do if... Oxi 1970i
50
ba75691e01 01/2007
Error message E7
E7E7
E7
AR flashes
continuously
Measured value too low
Measured value too
high
Display
to
toto
to
Sensor symbol flashes
Cause Remedy
Membrane damaged
Membrane damaged Regenerate sensor
Membrane head not screwed
on tight enough
Screw membrane head tight
Cause Remedy
No stable measured value
Membrane contaminated Clean membrane
Cause Remedy
Insufficient
flow
Provide flow to the sensor
Cause Remedy
High amount of dissolved
substances
Correct solubility function
using the salinity equivalent
Air bubbles bump on the
membrane with high velocity
Avoid direct flow to the
membrane
The carbon dioxide pressure is
too high (> 1 bar)
Measuring not possible
Cause Remedy
Time-out of the interface Check the instrument
connected
Cause Remedy
Calibration interval expired Recalibrate the measuring
system