Operating Manual System OxiTop® Control OxiTop® OC100 Controller OxiTop®-C Measuring Heads OC Model 0 BA31114e5 07/2006 BA31114e5 07/2006
Safety Instructions Please read these safety instructions carefully before installing the instrument! This instrument is built and checked according to IEC 1010, safety rules for electronic measuring instruments and left the factory secure from a safety engineering aspect. The smooth functioning and operational safety of the equipment can only be guaranteed by following the general safety precautions applicable and the special safety instructions given in this operating manual.
Safety Instructions Please read these safety instructions carefully before installing the instrument! This instrument is built and checked according to IEC 1010, safety rules for electronic measuring instruments and left the factory secure from a safety engineering aspect. The smooth functioning and operational safety of the equipment can only be guaranteed by following the general safety precautions applicable and the special safety instructions given in this operating manual.
Contents Contents Safety Instructions ....................................................................................................3 Contents.....................................................................................................................4 Introduction ...............................................................................................................6 The measuring principle ........................................................................................
Contents Contents Safety Instructions ....................................................................................................3 Contents.....................................................................................................................4 Introduction ...............................................................................................................6 The measuring principle ........................................................................................
Introduction Introduction ® In addition, in conjunction with the enhanced OxiTop Controller OC110 and relevant ® accessories, the OxiTop -C measuring heads can also be used to • Measure anaerobic degradation processes (e.g. biogas evaluation) The measuring principle The sample bottles are filled according to the data visible in the display (BOD measuring range with assigned water sample amount).
Introduction Introduction ® In addition, in conjunction with the enhanced OxiTop Controller OC110 and relevant ® accessories, the OxiTop -C measuring heads can also be used to • Measure anaerobic degradation processes (e.g. biogas evaluation) The measuring principle The sample bottles are filled according to the data visible in the display (BOD measuring range with assigned water sample amount).
General Operating Principles Switching On Controller keyboard Operating modes The OC100 controller has 2 standard operating modes that differ from one another as described below: ”Routine BOD” This mode provides the user-friendly ”quick introduction”. The samples are managed as single samples. The setting functions are logically restricted.
General Operating Principles Switching On Controller keyboard Operating modes The OC100 controller has 2 standard operating modes that differ from one another as described below: ”Routine BOD” This mode provides the user-friendly ”quick introduction”. The samples are managed as single samples. The setting functions are logically restricted.
General Operating Principles Switching On Switching on the controller Representation of keys and displays: Example: Pressing the ”Communication” button causes the mode to change to the ”Communication with the measuring heads” mode of operation: Required action: press button. The instrument displays: The current date and time appear (important for the allocation of sample numbers). If the date/time is incorrect, undertake corrections in ”GLP/Tools”.
General Operating Principles Switching On Switching on the controller Representation of keys and displays: Example: Pressing the ”Communication” button causes the mode to change to the ”Communication with the measuring heads” mode of operation: Required action: press button. The instrument displays: The current date and time appear (important for the allocation of sample numbers). If the date/time is incorrect, undertake corrections in ”GLP/Tools”.
Operating Mode: Routine BOD Operating Mode: Routine BOD Confirm the selected measurement range for the sample. The automatically allocated sample number is given in the header line (YY/MM/DD and sequential number). Operating mode: Routine BOD Sample preparation See WTW application reports (contained within the scope of delivery of the accessories supplied). Additional information: Type of measurement, run time, measurement range, final date, Id number.
Operating Mode: Routine BOD Operating Mode: Routine BOD Confirm the selected measurement range for the sample. The automatically allocated sample number is given in the header line (YY/MM/DD and sequential number). Operating mode: Routine BOD Sample preparation See WTW application reports (contained within the scope of delivery of the accessories supplied). Additional information: Type of measurement, run time, measurement range, final date, Id number.
Operating Mode: Routine BOD Displays “Started“. From this point in time, the sample exists within the sample management. Subsequently, the controller returns automatically to the input menu. Operating Mode: Routine BOD Call up all data This function is used to call up the data of all measuring heads. For the call up of the data of individual measuring heads: see the chapter on sample management. Entry menu “Communication with the measuring heads“. If no sample starts (e.g.
Operating Mode: Routine BOD Displays “Started“. From this point in time, the sample exists within the sample management. Subsequently, the controller returns automatically to the input menu. Operating Mode: Routine BOD Call up all data This function is used to call up the data of all measuring heads. For the call up of the data of individual measuring heads: see the chapter on sample management. Entry menu “Communication with the measuring heads“. If no sample starts (e.g.
Operating Mode: Routine BOD If all the measuring heads do not respond in the 1st pass of the data query, the controller searches for the missing measuring heads for approximately a further 7 seconds in the scanner mode. Without having to press another key, you can continue the data queries at other locations (e.g. other stirring platforms or other thermal cabinets).
Operating Mode: Routine BOD If all the measuring heads do not respond in the 1st pass of the data query, the controller searches for the missing measuring heads for approximately a further 7 seconds in the scanner mode. Without having to press another key, you can continue the data queries at other locations (e.g. other stirring platforms or other thermal cabinets).
Operating Mode: Routine BOD Select a sample. Operating Mode: Routine BOD Show sample This function is used to find the measuring heads. A sample labeling is not necessary. Depending on the sample selected, one of the two following menus appears. The header line contains the: - sample number - Id number (e.g. I001) - BOD type and run time of the selected sample. Finished sample Starting point: sample management . Use to select a sample. Current sample Use to select “Show sample“.
Operating Mode: Routine BOD Select a sample. Operating Mode: Routine BOD Show sample This function is used to find the measuring heads. A sample labeling is not necessary. Depending on the sample selected, one of the two following menus appears. The header line contains the: - sample number - Id number (e.g. I001) - BOD type and run time of the selected sample. Finished sample Starting point: sample management . Use to select a sample. Current sample Use to select “Show sample“.
Operating Mode: Routine BOD Operating Mode: Routine BOD Show measuring head list Erase data of finished samples This function erases the data of finished samples from the sample management of the controller. You can only erase a sample if it is finished; in non-finished samples, the menu item “Erase sample“ does not appear. This function is used to allocate a measuring head to the relevant sample. Use to select “ Starting point: sample management . list“. Use to select a finished sample.
Operating Mode: Routine BOD Operating Mode: Routine BOD Show measuring head list Erase data of finished samples This function erases the data of finished samples from the sample management of the controller. You can only erase a sample if it is finished; in non-finished samples, the menu item “Erase sample“ does not appear. This function is used to allocate a measuring head to the relevant sample. Use to select “ Starting point: sample management . list“. Use to select a finished sample.
Operating Mode: Routine BOD Operating Mode: Routine BOD The controller then returns automatically to the submenu. Call up data This function is used to call up the data of individual measuring heads. To call up the data of all measuring heads, see the chapter “Call up all data“.
Operating Mode: Routine BOD Operating Mode: Routine BOD The controller then returns automatically to the submenu. Call up data This function is used to call up the data of individual measuring heads. To call up the data of all measuring heads, see the chapter “Call up all data“.
Operating Mode: Routine BOD Call up data - Stop Operating Mode: Routine BOD Evaluation If a measuring head is missing or defective and the controller cannot transfer the data as a result, the menu item “Stop“ is provided. Starting point: sample management. Use to select a sample. After confirmation, the display message shown here appears. Enters into the evaluation: Display of the selected sample as a curve together with measured value data.
Operating Mode: Routine BOD Call up data - Stop Operating Mode: Routine BOD Evaluation If a measuring head is missing or defective and the controller cannot transfer the data as a result, the menu item “Stop“ is provided. Starting point: sample management. Use to select a sample. After confirmation, the display message shown here appears. Enters into the evaluation: Display of the selected sample as a curve together with measured value data.
Operating Mode: Routine BOD Operating Mode: Routine BOD The measured value undercuts the measurement range (underflow). Curve display of samples that are too cold Display of a curve The measured value temporarily leaves the measurement range. In using samples that are too cold, the maximum AutoTemp time is not sufficient to reach the temperature adaption of the sample: By warming the sample, excess pressure can result.
Operating Mode: Routine BOD Operating Mode: Routine BOD The measured value undercuts the measurement range (underflow). Curve display of samples that are too cold Display of a curve The measured value temporarily leaves the measurement range. In using samples that are too cold, the maximum AutoTemp time is not sufficient to reach the temperature adaption of the sample: By warming the sample, excess pressure can result.
Operating Mode: Routine BOD From the sample management: From the evaluation: 28 BA31114e5 07/2006 BA31114e5 07/2006 29
Operating Mode: Standard BOD Operating mode: Standard BOD Sample preparation See WTW application reports (contained within the scope of delivery of the accessories supplied). Screw the OxiTop -C measuring heads onto the sample bottles and close them tightly. Important note: Never use joint grease or other lubricant for your OxiTop-C measuring heads. Some of these products contain solvents that can cause severe damage to the plastic housing.
Operating Mode: Standard BOD Operating Mode: Standard BOD Confirm the selected measurement range for the sample. The automatically assigned sample number (YY/MM/DD and sequential number) is given in the header line. The controller repeatedly sends the start information in the scanner mode until successful feedback is received from the OxiTop®-C measuring head. After the successful start message from the measuring head: Displays “!Started!”.
Operating Mode: Standard BOD Operating Mode: Standard BOD Confirm the selected measurement range for the sample. The automatically assigned sample number (YY/MM/DD and sequential number) is given in the header line. The controller repeatedly sends the start information in the scanner mode until successful feedback is received from the OxiTop®-C measuring head. After the successful start message from the measuring head: Displays “!Started!”.
Operating Mode: Standard BOD Operating Mode: Standard BOD The start was stopped. Use to select “Call up all data”. to confirm Use start” and hold the “Continue controller to a measuring head (see above). Select “Stop start” to return to entry level menu. If an attempt is made to start a measuring head that was already started: Displays “Measuring head already used!” Point the controller at the measuring heads: to confirm “New Use ” and hold the controller to a free measuring head (see above).
Operating Mode: Standard BOD Operating Mode: Standard BOD The start was stopped. Use to select “Call up all data”. to confirm Use start” and hold the “Continue controller to a measuring head (see above). Select “Stop start” to return to entry level menu. If an attempt is made to start a measuring head that was already started: Displays “Measuring head already used!” Point the controller at the measuring heads: to confirm “New Use ” and hold the controller to a free measuring head (see above).
Operating Mode: Standard BOD Operating Mode: Standard BOD the chapter “What to do if...“ Point the controller at the measuring heads: Sample management 40 cm to 100 cm Entry into the sample management. A list of samples appears in the display (if samples are available): The call up of further measuring heads is performed in the scanner mode. Message that all the data of all the measuring heads has been called up. The instrument then returns to the input menu.
Operating Mode: Standard BOD Operating Mode: Standard BOD the chapter “What to do if...“ Point the controller at the measuring heads: Sample management 40 cm to 100 cm Entry into the sample management. A list of samples appears in the display (if samples are available): The call up of further measuring heads is performed in the scanner mode. Message that all the data of all the measuring heads has been called up. The instrument then returns to the input menu.
Operating Mode: Standard BOD Operating Mode: Standard BOD According to the sample selected, one of the two following menus appears. The header line contains the: - sample number - Id number (e.g. I001) - BOD type and run time of the selected sample. Finished sample Point the controller at the measuring heads: 40 cm to 100 cm The controller transmits the call up of the selected sample. The allocated measuring head flashes for approx. 5 seconds.
Operating Mode: Standard BOD Operating Mode: Standard BOD According to the sample selected, one of the two following menus appears. The header line contains the: - sample number - Id number (e.g. I001) - BOD type and run time of the selected sample. Finished sample Point the controller at the measuring heads: 40 cm to 100 cm The controller transmits the call up of the selected sample. The allocated measuring head flashes for approx. 5 seconds.
Operating Mode: Standard BOD Operating Mode: Standard BOD Safety prompt with possibility to return. The sample has been erased. List of measuring heads of the selected sample is displayed together with the serial numbers of the measuring heads. Point the controller at the measuring heads: 40 cm to 100 cm to select the measuring head and use to Use confirm. The controller again transmits the call up of the selected sample. The measuring head flashes for approx. 5 seconds.
Operating Mode: Standard BOD Operating Mode: Standard BOD Safety prompt with possibility to return. The sample has been erased. List of measuring heads of the selected sample is displayed together with the serial numbers of the measuring heads. Point the controller at the measuring heads: 40 cm to 100 cm to select the measuring head and use to Use confirm. The controller again transmits the call up of the selected sample. The measuring head flashes for approx. 5 seconds.
Operating Mode: Standard BOD Operating Mode: Standard BOD The controller then returns automatically to the submenu. Jumps to submenu of the selected sample.
Operating Mode: Standard BOD Operating Mode: Standard BOD The controller then returns automatically to the submenu. Jumps to submenu of the selected sample.
Operating Mode: Standard BOD Operating Mode: Standard BOD Call up data - Stop Erase: Removes missing measuring heads from the data stock. The completed sample is given the “finished” status. then If a measuring head is missing or defective and the controller cannot completely call up the sample as a result, the menu item “Stop“ is provided to stop the data transfer. Back: Jumps back to the previous menu. Abort using a function key: The stopped sample is not declared as “finished”.
Operating Mode: Standard BOD Operating Mode: Standard BOD Call up data - Stop Erase: Removes missing measuring heads from the data stock. The completed sample is given the “finished” status. then If a measuring head is missing or defective and the controller cannot completely call up the sample as a result, the menu item “Stop“ is provided to stop the data transfer. Back: Jumps back to the previous menu. Abort using a function key: The stopped sample is not declared as “finished”.
Operating Mode: Standard BOD Operating Mode: Standard BOD Evaluation Sample statistics Starting point: display of a single curve. Starting point: Sample management. Use sample. to select the Changes to sample statistics. Printout of the results with curves. The menu selection shown here only appears if n (the number of measuring heads) is at least 2. If n = 1, see the chapter “Routine BOD - Evaluation”. Evaluation of the whole sample: Display of all the curves together with mean value data.
Operating Mode: Standard BOD Operating Mode: Standard BOD Evaluation Sample statistics Starting point: display of a single curve. Starting point: Sample management. Use sample. to select the Changes to sample statistics. Printout of the results with curves. The menu selection shown here only appears if n (the number of measuring heads) is at least 2. If n = 1, see the chapter “Routine BOD - Evaluation”. Evaluation of the whole sample: Display of all the curves together with mean value data.
Operating Mode: Standard BOD Operating Mode: Standard BOD Message “Curve excluded“. Excluding a curve This function is used to temporarily exclude a single curve (e.g. an outlier) from the evaluation and averaging of a whole sample. The curve is only excluded temporarily! The data stock of the sample management does not change. The excluded curve is present again when the call is repeated. Updated display (curve excluded, mean value newly calculated).
Operating Mode: Standard BOD Operating Mode: Standard BOD Message “Curve excluded“. Excluding a curve This function is used to temporarily exclude a single curve (e.g. an outlier) from the evaluation and averaging of a whole sample. The curve is only excluded temporarily! The data stock of the sample management does not change. The excluded curve is present again when the call is repeated. Updated display (curve excluded, mean value newly calculated).
Operating Mode: Standard BOD Operating Mode: Standard BOD Curves display for cold samples Cursor interrogation Starting point: Display of a single curve with measured value data. Display of a single curve. Use to change to “Cursor query“. Display of all curves. Use to run through the curve values. When using samples that are too cold, the maximum AutoTemp time is not sufficient to reach the temperature balance of the sample: Warming the sample can cause excess pressure.
Operating Mode: Standard BOD Operating Mode: Standard BOD Curves display for cold samples Cursor interrogation Starting point: Display of a single curve with measured value data. Display of a single curve. Use to change to “Cursor query“. Display of all curves. Use to run through the curve values. When using samples that are too cold, the maximum AutoTemp time is not sufficient to reach the temperature balance of the sample: Warming the sample can cause excess pressure.
Operating Mode: Standard BOD Operating Mode: Standard BOD Print Measured values outside the measuring range In the following cases, “undef.“ (undefined) is displayed instead of the measured value or mean value during evaluation: - A measured curve exceeding the measuring range at any point of its path. - A measured curve undercutting the measuring range at its end point. Switch on the TD100 IR printer.
Operating Mode: Standard BOD Operating Mode: Standard BOD Print Measured values outside the measuring range In the following cases, “undef.“ (undefined) is displayed instead of the measured value or mean value during evaluation: - A measured curve exceeding the measuring range at any point of its path. - A measured curve undercutting the measuring range at its end point. Switch on the TD100 IR printer.
Operating Mode: Standard BOD Print from the evaluation: If the controller is located - in the display of all the curves with cursor display or - in the display of single curves with or without cursor query, the printout shows a copy of the relevant display.
GLP / Tools GLP/Tools main menu Main menu: Use to select a submenu Use to change to the submenu Submenus of the GLP/Configuration: Show free This function is used to identify free measuring heads without requiring sample labeling. Free measuring heads can be used to start new samples. Show settings The current settings are displayed here.
GLP / Tools Settings GLP / Tools Here you can undertake or change the following settings: Show settings Operating mode Measuring time Date/time * GLP; calibrating interval Memory * AutoTemp * Switch-off interval Language Check Maintenance Main menu, “GLP/Tools“. Use to select the “Show settings“ menu. (finished/all) Show info (with report printout) Controller info (with report printout) Cal test Pneumatic test The following example illustrates the presettings in the Standard BOD operating mode.
GLP / Tools Settings GLP / Tools Here you can undertake or change the following settings: Show settings Operating mode Measuring time Date/time * GLP; calibrating interval Memory * AutoTemp * Switch-off interval Language Check Maintenance Main menu, “GLP/Tools“. Use to select the “Show settings“ menu. (finished/all) Show info (with report printout) Controller info (with report printout) Cal test Pneumatic test The following example illustrates the presettings in the Standard BOD operating mode.
GLP / Tools GLP / Tools Settings Measuring time Set the type and measuring time of the BOD measurement here. Operating mode “Settings“ submenu: Main menu, “GLP/Tools“. Use to select the “Measuring time“ submenu. Use to select the “Settings“ menu. “Settings“ submenu: Use to set the days (1 to 99) or hours (0.5 to 23). Use to select “Operating mode“. Confirm using . Display: preset measuring time. Default on delivery: 5 days.
GLP / Tools GLP / Tools Settings Measuring time Set the type and measuring time of the BOD measurement here. Operating mode “Settings“ submenu: Main menu, “GLP/Tools“. Use to select the “Measuring time“ submenu. Use to select the “Settings“ menu. “Settings“ submenu: Use to set the days (1 to 99) or hours (0.5 to 23). Use to select “Operating mode“. Confirm using . Display: preset measuring time. Default on delivery: 5 days.
GLP / Tools GLP / Tools Switch the GLP function on or off (only in the standard BOD operating mode): Set each of the number blocks underlaid in black using “Settings“ submenu: . Confirm and continue with Use to select the “GLP“ submenu. . Set the day, month, year, hours, minutes, seconds consecutively. The measuring time of samples already started is not affected by a change of date and time. “GLP“ submenu. “GLP“ is preselected.
GLP / Tools GLP / Tools Switch the GLP function on or off (only in the standard BOD operating mode): Set each of the number blocks underlaid in black using “Settings“ submenu: . Confirm and continue with Use to select the “GLP“ submenu. . Set the day, month, year, hours, minutes, seconds consecutively. The measuring time of samples already started is not affected by a change of date and time. “GLP“ submenu. “GLP“ is preselected.
GLP / Tools GLP / Tools Press , then set the time period for the calibrating interval (1 - 36 months) using Default: 12 months. . Confirm by . Memory Here, you can set whether the controller should automatically erase the oldest finished samples if the memory is full to create space for new measurement data (setting “auto“). With the “manual“ setting, the message “Lack of memory! Erase finished samples!“ is displayed if the memory is full (See chapter “What to do if ...?).
GLP / Tools GLP / Tools Press , then set the time period for the calibrating interval (1 - 36 months) using Default: 12 months. . Confirm by . Memory Here, you can set whether the controller should automatically erase the oldest finished samples if the memory is full to create space for new measurement data (setting “auto“). With the “manual“ setting, the message “Lack of memory! Erase finished samples!“ is displayed if the memory is full (See chapter “What to do if ...?).
GLP / Tools GLP / Tools Sequence of the test phase: With a further drop in pressure (consumption) or constant pressure after the adaption phase, the pressure value at the end of the adaption phase is the starting point of the measurement. On a further increase of pressure following the adaption phase (the sample is still too cold), the turning point of the pressure process at which the pressure increase changes into a pressure drop, is the starting point of the measurement.
GLP / Tools GLP / Tools Sequence of the test phase: With a further drop in pressure (consumption) or constant pressure after the adaption phase, the pressure value at the end of the adaption phase is the starting point of the measurement. On a further increase of pressure following the adaption phase (the sample is still too cold), the turning point of the pressure process at which the pressure increase changes into a pressure drop, is the starting point of the measurement.
GLP / Tools GLP / Tools Check Language Main menu, “GLP/Tools“. Select the language here in which the displays of the OxiTop® Controller appear. The controller has the following 5 languages stored in it (default German): German English - French - Italian - Spanish. Use to select the “Check“ menu. “Settings” menu: Use to select the “Language“ submenu. Show “Check“ menu. “Show “ preselected. The default language is “German“.
GLP / Tools GLP / Tools Check Language Main menu, “GLP/Tools“. Select the language here in which the displays of the OxiTop® Controller appear. The controller has the following 5 languages stored in it (default German): German English - French - Italian - Spanish. Use to select the “Check“ menu. “Settings” menu: Use to select the “Language“ submenu. Show “Check“ menu. “Show “ preselected. The default language is “German“.
GLP / Tools GLP / Tools If the measuring head does not respond, the following display appears after approximately 7 seconds: info “Check“ menu. Use info“. The query was stopped. You can select between - continue query (see above) and - Stop (return to the “Check“ menu to select “ Controller info “Check“ menu. Use to select “Controller info“.
GLP / Tools GLP / Tools If the measuring head does not respond, the following display appears after approximately 7 seconds: info “Check“ menu. Use info“. The query was stopped. You can select between - continue query (see above) and - Stop (return to the “Check“ menu to select “ Controller info “Check“ menu. Use to select “Controller info“.
GLP / Tools GLP / Tools Evaluation of the Cal test Cal test Starting point: sample management. The Cal test is used to test the sealing of the system measuring head - rubber sleeve - sample bottle and the operability of the systems OxiTop® Control. In the standard BOD operating mode with the “GLP ON” setting, the controller indicates when the next Cal test is due after each set up calibration interval (see the chapter “GLP/Tools - GLP”),.
GLP / Tools GLP / Tools Evaluation of the Cal test Cal test Starting point: sample management. The Cal test is used to test the sealing of the system measuring head - rubber sleeve - sample bottle and the operability of the systems OxiTop® Control. In the standard BOD operating mode with the “GLP ON” setting, the controller indicates when the next Cal test is due after each set up calibration interval (see the chapter “GLP/Tools - GLP”),.
GLP / Tools GLP / Tools Pneumatic test The controller sets a new test date in the measuring head. The controller calculates the new test date from the current date + the check interval set up (see the chapter “GLP/Tools - Settings GLP - Check interval”). The pneumatic test tests the measurement precision of the measuring head. It says nothing about the long-term impermeability of the system. To perform the pneumatic test, you require the testing agent, OxiTop® PT (WTW order number 209 334).
GLP / Tools GLP / Tools Pneumatic test The controller sets a new test date in the measuring head. The controller calculates the new test date from the current date + the check interval set up (see the chapter “GLP/Tools - Settings GLP - Check interval”). The pneumatic test tests the measurement precision of the measuring head. It says nothing about the long-term impermeability of the system. To perform the pneumatic test, you require the testing agent, OxiTop® PT (WTW order number 209 334).
GLP / Tools GLP / Tools A submenu appears with the selection: - From sample no. (preselected) - All - Back (from the “Maintenance” submenu) (Exceeding the time causes a return to the “Check“ menu). The controller display shows the result of the pneumatic test. Use “Check“ menu. The controller displays the list of finished samples. The oldest finished sample is marked. to return to the With , you can mark further samples. Maintenance , you can remove the With marking again. Main menu, “GLP/Tools“.
GLP / Tools GLP / Tools A submenu appears with the selection: - From sample no. (preselected) - All - Back (from the “Maintenance” submenu) (Exceeding the time causes a return to the “Check“ menu). The controller display shows the result of the pneumatic test. Use “Check“ menu. The controller displays the list of finished samples. The oldest finished sample is marked. to return to the With , you can mark further samples. Maintenance , you can remove the With marking again. Main menu, “GLP/Tools“.
GLP / Tools GLP / Tools In the selection of erase “All“ samples, the following display appears: The controller asks again if you really want to erase the samples from the memory. Further: See above. The serial number of the measuring head and the sample number appear on the display. If no finished samples are available in the memory, the following display appears: if you want to Press release the measuring head. (If you do not want to release the measuring head, select and confirm it with “Return“.
GLP / Tools GLP / Tools In the selection of erase “All“ samples, the following display appears: The controller asks again if you really want to erase the samples from the memory. Further: See above. The serial number of the measuring head and the sample number appear on the display. If no finished samples are available in the memory, the following display appears: if you want to Press release the measuring head. (If you do not want to release the measuring head, select and confirm it with “Return“.
RS232 Interface Cleaning RS232 interface Cleaning the sample bottles See the WTW application report. Cleaning the controller and measuring heads RS232 interface • Do not use any solvent (such as alcohol or acetone)! The RS232 interface is used for communication with the computer (PC). The RS232 interface is provided solely for the implementation of the WTW software, “Achat OC“ (WTW software “ACHAT OC“ and RS232 interface cable AK 540/B available from 1998).
RS232 Interface Cleaning RS232 interface Cleaning the sample bottles See the WTW application report. Cleaning the controller and measuring heads RS232 interface • Do not use any solvent (such as alcohol or acetone)! The RS232 interface is used for communication with the computer (PC). The RS232 interface is provided solely for the implementation of the WTW software, “Achat OC“ (WTW software “ACHAT OC“ and RS232 interface cable AK 540/B available from 1998).
Power Supply Power Supply The OxiTop®-OC100 and OxiTop®-C instruments are battery-powered. To ensure reliable operation, both instruments have a 2-stage battery status monitor 1st level = warning level : Batteries LoBat ! 2nd level = error level : Batteries empty ! Changing supply batteries: • • • • OxiTop®-OC100 controller Economy circuit (automatic switch off) The instrument switches off automatically following the last key actuation after expiry of the set switch-off interval.
Power Supply Power Supply The OxiTop®-OC100 and OxiTop®-C instruments are battery-powered. To ensure reliable operation, both instruments have a 2-stage battery status monitor 1st level = warning level : Batteries LoBat ! 2nd level = error level : Batteries empty ! Changing supply batteries: • • • • OxiTop®-OC100 controller Economy circuit (automatic switch off) The instrument switches off automatically following the last key actuation after expiry of the set switch-off interval.
Power Supply Power Supply Changing the data backup battery: • • • • • • • • Evaluate and save all the measurement data. Options available : - print out the results, see the chapter on printing - save the data in the PC using the software “Achat OC“, see the chapter “RS232 interface” - handwritten documentation of the results. Documentation of your instrument settings: - printout of your instrument settings, see the chapter “GLP/Tools - Settings Show settings”. Switch off the OxiTop®-OC100 controller.
Power Supply Power Supply Changing the data backup battery: • • • • • • • • Evaluate and save all the measurement data. Options available : - print out the results, see the chapter on printing - save the data in the PC using the software “Achat OC“, see the chapter “RS232 interface” - handwritten documentation of the results. Documentation of your instrument settings: - printout of your instrument settings, see the chapter “GLP/Tools - Settings Show settings”. Switch off the OxiTop®-OC100 controller.
Power Supply Power Supply Changing the supply battery OxiTop®-C measuring head • Change the battery after the measurement has been terminated. Supply batteries Batteries: 2 x lithium batteries, CR2430, e.g. WTW type Batt/OxiTop, order no. 209 012 These batteries ensure the energy supply of the OxiTop®-C measuring head. Run time: typically 2 years The OxiTop®-C measuring head does not require any batteries for data backup.
Power Supply Power Supply Changing the supply battery OxiTop®-C measuring head • Change the battery after the measurement has been terminated. Supply batteries Batteries: 2 x lithium batteries, CR2430, e.g. WTW type Batt/OxiTop, order no. 209 012 These batteries ensure the energy supply of the OxiTop®-C measuring head. Run time: typically 2 years The OxiTop®-C measuring head does not require any batteries for data backup.
What to do if...? What to do if...? Display messages Messages already used ! Set system clock ! defective ! No samples available ! 84 Messages Explanations, causes and problem solutions No active Explanations, causes and problem solutions Causes: • No measurement has been started. • The measuring heads are placed at another location (e.g. other levels in the incubator, other incubators, further storage locations).
What to do if...? What to do if...? Display messages Messages already used ! Set system clock ! defective ! No samples available ! 84 Messages Explanations, causes and problem solutions No active Explanations, causes and problem solutions Causes: • No measurement has been started. • The measuring heads are placed at another location (e.g. other levels in the incubator, other incubators, further storage locations).
What to do if...? What to do if...? Messages Explanations, causes and problem solutions Messages Explanations, causes and problem solutions Lack of memory possible ! Cause: This message is used as information in the standard BOD operating mode. There is a lot of sample data stored in the sample management. The memory is almost full. It is still possible to start a sample as a parallel sample start with 11 measuring heads (the number 11 used here is an example, possible numbers: 1...11 ).
What to do if...? What to do if...? Messages Explanations, causes and problem solutions Messages Explanations, causes and problem solutions Lack of memory possible ! Cause: This message is used as information in the standard BOD operating mode. There is a lot of sample data stored in the sample management. The memory is almost full. It is still possible to start a sample as a parallel sample start with 11 measuring heads (the number 11 used here is an example, possible numbers: 1...11 ).
What to do if...? What to do if...? Messages Explanations, causes and problem solutions 10 of 12 Cause: • Two measuring heads have not been found by the controller • The measuring heads are positioned at another location, e.g. other levels in the incubator, other incubators, further storage positions • The measuring heads have no optical contact with the controller: - Distance too great, angular position imprecise - IR window of the measuring head averted from the controller - Shading by other objects.
What to do if...? What to do if...? Messages Explanations, causes and problem solutions 10 of 12 Cause: • Two measuring heads have not been found by the controller • The measuring heads are positioned at another location, e.g. other levels in the incubator, other incubators, further storage positions • The measuring heads have no optical contact with the controller: - Distance too great, angular position imprecise - IR window of the measuring head averted from the controller - Shading by other objects.
What to do if...? What to do if...? Requirements / Problems Procedure / Problem solutions Requirements / Problems Procedure / Problem solutions Measuring head unintentionally started for measurement Problem solution The measuring head can be released again through the command Reset/release: Chapter “GLP/Tools - Maintenance - Reset/release “. The controller guides you further through the operation.
What to do if...? What to do if...? Requirements / Problems Procedure / Problem solutions Requirements / Problems Procedure / Problem solutions Measuring head unintentionally started for measurement Problem solution The measuring head can be released again through the command Reset/release: Chapter “GLP/Tools - Maintenance - Reset/release “. The controller guides you further through the operation.
What to do if...? Requirements / Problems What to do if...? Procedure / Problem solutions Requirements / Problems Checking the controller Problem solution: Controller info • See the chapter “GLP/Tools - Check - Controller info” All single information on the state of the controller is listed. • The check can be undertaken at any time without affecting the measurement that is running. • Perform the command: “Show all “. See “GLP/Tools - Check - Show ”.
What to do if...? Requirements / Problems What to do if...? Procedure / Problem solutions Requirements / Problems Checking the controller Problem solution: Controller info • See the chapter “GLP/Tools - Check - Controller info” All single information on the state of the controller is listed. • The check can be undertaken at any time without affecting the measurement that is running. • Perform the command: “Show all “. See “GLP/Tools - Check - Show ”.
Technical Data Accessories and spare parts OxiTop® OC 100 controller Accessories Measurement ranges Run times and data sets See WTW General Catalog and WTW application reports. Measurement period Spare parts Description Type Order no.
Technical Data Accessories and spare parts OxiTop® OC 100 controller Accessories Measurement ranges Run times and data sets See WTW General Catalog and WTW application reports. Measurement period Spare parts Description Type Order no.
Technical Data Technical Data OxiTop®-C measuring head Measuring principle manometric by means of a pressure measuring head Measured of BODn (oxygen drift according to DIN 38409 T 52) Pressure range 500 ... 1350 hPa (mbar). Accuracy ± 1% of measured value ± 0.7 % of the measuring range Resolution 1 hPa Display LED pilot lamps Power supply 2 x lithium batteries, CR2430, e.g. WTW type Batt/OxiTop, order no.
Technical Data Technical Data OxiTop®-C measuring head Measuring principle manometric by means of a pressure measuring head Measured of BODn (oxygen drift according to DIN 38409 T 52) Pressure range 500 ... 1350 hPa (mbar). Accuracy ± 1% of measured value ± 0.7 % of the measuring range Resolution 1 hPa Display LED pilot lamps Power supply 2 x lithium batteries, CR2430, e.g. WTW type Batt/OxiTop, order no.
Supplement to the operating manual System OxiTop® Control Controller ® OxiTop OC100 Measuring heads OxiTop®-C This sheet is a supplement to the present operating manual and contains the chapter “Restore data“ and some notes. Always keep this sheet together with the operating manual.
Supplement to the operating manual System OxiTop® Control Controller ® OxiTop OC100 Measuring heads OxiTop®-C This sheet is a supplement to the present operating manual and contains the chapter “Restore data“ and some notes. Always keep this sheet together with the operating manual.
Supplement to the operating manual Supplement to the operating manual After reconfirming the data of the measuring head is read out and stored in the controller. Restore data In case you lost your controller or your controller is defective, the “Restore data“ function facilitates to restore the data of running measurements using a new or another controller. This means the data are not lost. The Id number of the sample is not restored, it is always I999.
Supplement to the operating manual Supplement to the operating manual After reconfirming the data of the measuring head is read out and stored in the controller. Restore data In case you lost your controller or your controller is defective, the “Restore data“ function facilitates to restore the data of running measurements using a new or another controller. This means the data are not lost. The Id number of the sample is not restored, it is always I999.
Supplement to the operating manual Supplement to the operating manual Notes • • • 4 The present operating manual and this supplement refer to the software release 2.xx. The right to implement minor changes is reserved. The OxiTop Control 100 measuring system can be extended up to 120 measuring positions (measuring heads) – in the operating manual there are 144 measuring positions mentioned. The OC100 controller reliably saves your measuring data. However, please back up your data regularly as usual.
Fax: E-Mail: Internet: Tel: +49 (0) 881 183-0 +49 (0) 881 183-100 +49 (0) 881 183-420 Info@WTW.com http://www.WTW.com Germany Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Wissenschaftlich-Technische Werkstätten GmbH Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: Fax: E-Mail: Internet: +49 (0) 881 183-0 +49 (0) 881 183-100 +49 (0) 881 183-420 Info@WTW.com http://www.WTW.