Owner manual

Power Supply
Power Supply
78 BA31114e5 07/2006 BA31114e5 07/2006 79
The OxiTop
®
-OC100 and OxiTop
®
-C instruments are battery-powered.
To ensure reliable operation, both instruments have a 2-stage battery status monitor
1st level = warning level : Batteries LoBat !
2nd level = error level : Batteries empty !
OxiTop
®
-OC100 controller
Economy circuit (automatic switch off)
The instrument switches off automatically following the last key actuation after expiry
of the set switch-off interval.
Switch-off interval:
Routine BOD operating mode: preset to a fixed period of 5 min.
Standard BOD operating mode: delivery state of 5 min.,
settings of 5...15 min. possible
Supply batteries
Batteries: 3 pcs, alkaline (alkaline manganese), size: Mignon, AA, AM3, LR6
These batteries ensure the energy supply of the OxiTop
®
OC100.
Run time: > 100 h (approx. 1000 start-ups in normal use)
Supply batteries status signals
are given on switching on the instrument by display messages and a signal tone
or can be called up under “GLP/Tools - Check - Controller info”.
Display message
Note / Meaning
Supply
Battery/ies LoBat !
Please change !
The warning appears for approx. 3 seconds.
The instrument then continues to run normally.
The instrument can still be safely operated within the
specifications.
When the message first appears, there is still a
running reserve available.
Please obtain new batteries and replace the old ones
!
Supply
Battery/ies empty!
Please change !
The message appears for approx. 3 seconds.
The instrument then switches itself off.
The instrument can no longer be used.
The supply batteries must be replaced by new ones.
Changing supply batteries:
Switch off the OxiTop
®
-OC100 controller.
Loosen the 4 screws underneath the housing using a Phillips screwdriver (see
figure
).
Place the controller on the lower case.
Remove the upper case and put it down to the right next to the lower case with
the display downwards (see figure
).
Remove the supply battery holder from the fixing in the lower case and turn it
around.
Remove the empty supply batteries.
Insert new supply batteries - 3 pieces, alkaline (alkaline manganese),
size: Mignon, AA, AM3, LR6).
Ensure that they are the right way round!
(The poles are marked in the supply battery holder)
Always replace the complete set of batteries.
Turn the supply battery holder around again and place it in the fixing in the lower
case.
Set the upper case on top of the lower case, turn the controller and tighten the
housing screws using the screwdriver (see figure
).
Switch on the controller. The battery change was successful if no error message
appears concerning the supply batteries.
It is recommended to use only brand name batteries of the type specified as
“Alkaline“. Using other types of battery can affect reliable functioning.
The supply batteries have no influence on data integrity.