Owner manual

Power Supply
Power Supply
80 BA31114e5 07/2006 BA31114e5 07/2006 81
Data backup battery
1 x lithium battery, CR2430, e.g. WTW type Batt/OxiTop
,
order no. 209 012. This battery ensures that your measurement data and the
instrument settings in the OxiTop
®
- OC100 controller are saved. Run time: typically 4
years
Data backup battery - status messages
are given on switching on the instrument by display messages and a signal tone
or can be called up under “GLP/Tools - Check - Controller info”.
Display message
Note / Meaning
Data protection battery/ies LoBat!
Please change !
The warning appears for approx. 3 seconds.
The instrument then continues to run normally.
The instrument can still be operated within the
specifications.
When the message first appears, there is still a
running reserve available. Please obtain new
batteries and replace the old ones.
Changing the data backup battery:
Evaluate and save all the measurement data.
Options available :
- print out the results, see the chapter on printing
- save the data in the PC using the software “Achat OC“, see the chapter
“RS232 interface”
- handwritten documentation of the results.
Documentation of your instrument settings:
- printout of your instrument settings, see the chapter “GLP/Tools - Settings -
Show settings”.
Switch off the OxiTop
®
-OC100 controller.
Loosen the 4 screws underneath the housing using a Phillips screwdriver (see
figure
).
Place the controller on the lower case.
Remove the upper case and put it down to the right next to the lower case with
the display downwards (see figure
).
Remove the supply battery holder from the fixing in the lower case and turn it
around.
Remove the empty data backup battery.
Insert the new data backup battery (1 piece, lithium battery, CR2430,
e.g. WTW type Batt/OxiTop
, order no. 209 012)
Ensure that the battery is the right way round!
The positive pole must be at the top.
Set the upper case on top of the lower case, turn the controller and tighten the
housing screws using the screwdriver (see figure
).
Switch on the controller. The message “Please set the system clock!“ appears.
This message is a reminder and should be acknowledged by .
The battery change was successful if no error message appears concerning the
data backup battery.
Enter the current date/time and your settings.