Owner manual

Technical Data
Technical Data
96 BA31114e5 07/2006 BA31114e5 07/2006 97
OxiTop
®
-C measuring head
Measuring principle manometric by means of a pressure measuring head
Measured of BOD
n
(oxygen drift according to DIN 38409 T 52)
Pressure range
500 ... 1350 hPa (mbar).
Accuracy ± 1% of measured value ± 0.7 % of the measuring
range
Resolution 1 hPa
Display LED pilot lamps
Power supply 2 x lithium batteries, CR2430, e.g. WTW type
Batt/OxiTop
, order no. 209 012)
Battery run time Typically 2 years, in case of normal use
Safety class 3, equipment safety: IEC 1010, EN 61010 part 1
Protection type IP 54 DIN 40050
EMC
-Interference emission
-interference immunity
EN 50081-1, FCC Class A
EN 50082-1
Climate class 2, VDI/VDE 3540
Ambient temperature Storage: -25°C ... +65°C
Operation: +5°C ... +50°C
Dimensions H: 70 mm,
70 mm
Weight 95 g
Test certificates CE, TÜV, UL, CUL(UL/CSA)
Warranty 1 year