Owner's manual

Operation pH/ 3400i
24
strument uses the Nernst slope (-59,16 mV/pH at 25 °C) and deter-
mines the asymmetry of the electrode.
6 Thoroughly rinse the electrode with deionized water.
7 If required, enter the temperature of the second buffer solution
with <> <> while keeping the <RUN/ENTER> key de-
pressed.
8 Immerse the pH electrode in the second buffer solution.
9 Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
10 When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The probe symbol
shows the evaluation of the current calibration.
11 Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
U
AutoCal TECAutoCal TEC
mV
U
mV/pH
mol/L
8
Auto
Store
TempError
RCL
LoBat
CalError
ARng
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
%
°F
S
mg/L
% ppm
ISE
8
4
2
°C
TP
mV/pH
mol/L
8
Auto
Store
TempError
RCL
LoBat
CalError
ARng
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
%
°F
S
mg/L
% ppm
ISE
8
4
2
8
0
°C
TP
80
4
pH
mV
AR
0
AR
AutoCal DIN AutoCal DIN
pH
1 4
6
AutoCal DIN
Ident
ConCal
Auto
Store
ARng
0
L
s
TP
°C
AR
mol/L
8
RCL
LoBat
CalError
Time
Baud
Day.Month No.
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
pH
45
9
mV/pH
AutoCal TEC