Owner's manual
Operation pH/ 3400i
30
8 Thoroughly rinse the electrode with deionized water.
9 If required, enter the temperature of the second buffer solution
with <▲> <▼> while keeping the <RUN/ENTER> key de-
pressed.
10 Immerse the pH electrode in the second buffer solution.
11 Press the <RUN/ENTER> key.
12 Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
13 When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
14 Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
AutoCal DIN
AutoCal TEC