6.2 ISE system messages
Error message 0FL
0FL0FL
0FL
Error message E3
E3E3
E3
Concentration
display flashes
Cause Remedy
Ion-selective electrode:
Not connected Connect the electrode
Cable broken Replace the electrode
Cause Remedy
Ion-selective electrode:
Slope not in the range
50 ... 70 mV or
23 ... 35 mV
Replace the electrode
recalibrate
Moisture in the plug Dry plug
Electrode obsolete Replace the electrode
Electrode broken Replace the electrode
Measuring instrument:
Calibration procedures:
Incorrect order of the
standards with 3-point
calibration
Select the correct order
Socket damp Dry socket
Cause Remedy
In measurement:
The measured value is outside
the range determined through
calibration
Select the calibration
standards so that the
measured value is in between
them
In calibration:
Calibration standards lie too
closely together. Select other
calibration standards
U2 U1 5mV
U3 U2 5mV