Owner's manual

What to do if... pH/ 3400i
68
TP and Temp Error flash
Calibration Std 2 flashes
Calibration Std 3 flashes
6.3 General errors
Display LoBat
Instrument does not
react
to keystroke
Cause Remedy
In calibration:
Calibration standards do not
have the correct temperature
(max. temperature difference
±
2 °C)
Temper the calibration
standards
In measurement:
Test sample does not have the
correct temperature (max.
temperature difference
± 2 °C)
Temper the test sample
Cause Remedy
Standard 2 = Standard 1 Use different calibration
standards
Cause Remedy
Requirement not met:
Standard 3 > Standard 2, if Std
2 > Std 1 or
Standard 3 < Standard 2, if Std
2 < Std 1
Use different standard 3
Cause Remedy
Batteries almost empty Exchange the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Switch the instrument on
while pressing the <CAL> key