Manual
Operation pH 3110
30
ba75791e02 11/2008
4.5.2 Measurement settings
These settings apply to calibration and measurement (the default
setting is printed in bold).
Display during
calibration
(bUFF)
Display during calibration (BUFFER) pH (buffer nominal value),
U (combination electrode
voltage)
Unit of the value for the slope (SLO) mV/pH, %
Unit of the value for the zero point
(ASY)
mV, pH
Measured value resolution, pH 0.000 , 0.00 , 0.0
Measured value resolution, U 0.0 , 0
Temperature unit (UnIt) °C, °F
Calibration interval (Int.C) 0 ... 7 ... 999 d
1 With <M_> open the menu for measurement and calibration
settings.
The first setting is displayed.
2 Using <▲><▼>, select the pH or U display during calibration.
3 Confirm with <ENTER>.
SLO, the unit of the value for the slope (mV/pH or %) is
displayed.
H
p
F
F
U
b
pH
AutoCal DIN
AutoCal TEC