Manual

pH 3110 What to do if...
39
ba75791e02 11/2008
6 What to do if...
Error message
OFL, UFL
Error message
CalError
Cause Remedy
pH combination electrode:
Measured value outside the
measuring range
Use suitable combination
electrode
Air bubble in front of the
diaphragm
Remove air bubble
Air in the diaphragm Extract air or moisten
diaphragm
Cable broken Replace combination
electrode
Gel electrolyte dried out Replace combination
electrode
Cause Remedy
pH combination electrode:
The values determined for zero
point and slope of the
combination electrode are
outside the allowed limits.
Recalibrate
Diaphragm contaminated Clean diaphragm
Combination electrode broken Replace combination
electrode
Buffer solutions
Incorrect buffer solutions Change calibration
procedure
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions