ba75793e02 12/2008
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
Measured
parameter
Time interval Stability in the time interval
pH value 15 seconds
Δ pH: better than 0.01