pH 3210 What to do if...
57
ba75793e02 12/2008
6 What to do if...
Error message
OFL, UFL
Error message
Error
Cause Remedy
pH combination electrode:
Measured value outside the
measuring range
Use suitable combination
electrode
Air bubble in front of the
junction
Remove air bubble
Air in the junction Extract air or moisten
junction
Cable broken Replace combination
electrode
Gel electrolyte dried out Replace combination
electrode
Cause Remedy
pH combination electrode:
The values determined for zero
point and slope of the
combination electrode are
outside the allowed limits.
Recalibrate
Junction contaminated Clean junction
Combination electrode broken Replace combination
electrode
Buffer solutions:
Incorrect buffer solutions Change calibration
procedure
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions