User Manual

Operation pHotoFlex / pHotoFlex Turb
74
ba75491d04 07/2006
Output format, CSV
Data Description
1 Date Date of storing
2 Time Time of storing
3 ID adjusted ID
4 Program number only for the measured parameter,
Concentration
5 Citation form only for the measured parameter,
Concentration
6 AR/
BV
! AutoRead (measuring mode, pH & ORP)
! Blank value (measured parameter,
Concentration)
7 Dilution only for the measured parameter,
Concentration
8 Measured value ! Measured value or
! Upper/lower measuring range limit
(only with measured value status, OFL/
UFL)
9 Unit of 8 ! Unit of the measured value or
! Designation of dimensionless measured
values, e.g. <pH>
10 Measured value
status of 8
! VALID: Measured value valid
! INVALID: Measured value invalid
! UFL: Measured value below the lower
measuring range limit
! OFL: Measured value above the upper
measuring range limit
11 Temperature val-
ue / wavelength
of the measure-
ment
! Temperature (measuring mode, pH &
ORP)
! Wavelength (measured parameter,
Absorbance / % Transmission)
15.01.05;15:06:49;1;8;O2;[BW];[9 + 1];6.5;mmol/l;VALID;
15.01.05;15:09:52;0;;;[AR];;7.00;<pH>;VALID;25;°C;VALID
11 12 13
123456789
10