User Manual
pHotoFlex / pHotoFlex Turb What to do if...
87
ba75491d04 07/2006
6 What to do if...
6.1 General errors
Display, LoBat
Instrument does not
react to keystroke
RS232 interface
does not react
Error message,
Error
0, 8, 16, 16384
Cause Remedy
– The batteries or accumulator
pack are largely depleted
– Insert new batteries
– Charge the accumulator pack
(see section 3.2)
Cause Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <START/ENTER>
and <PRT> key
simultaneously.
Cause Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <START/ENTER>
and <PRT> key
simultaneously.
Cause Remedy
– Instrument error – Repeat measurement
– Meter defective,
send meter to WTW for repair
and quote the error number