What to do if... pHotoFlex / pHotoFlex Turb
88
ba75491d04 07/2006
6.2 Photometry
Measuring range
undercut or exceeded
Obviously incorrect
measured values
6.3 pH value / ORP voltage
Measuring range
undercut or exceeded
Cause Remedy
Program not suitable Select program with suitable
measuring range
Dilute the sample
Cause Remedy
Measurement disturbed by
external light
Close the external light cover.
Cell not correctly inserted Insert the cell so that it is
positioned on the bottom of
the cell shaft.
Cell contaminated Clean the cell
Cell shaft contaminated Clean the cell shaft
Dilution set incorrectly Set the dilution
Selected program unsuitable Select other program
Zero measurement incorrect Perform zero measurement
Blank value incorrect Remeasure the blank value
Cause Remedy
Electrode:
Air bubble in front of the
diaphragm
Remove air bubble
Air in the diaphragm Extract air or
moisten diaphragm
Gel electrolyte dried out Replace electrode
Test sample
The pH value lies outside the
measuring range
not possible