pHotoFlex / pHotoFlex Turb What to do if...
89
ba75491d04 07/2006
Measured value display
----
(calibration error)
No stable measured
value
Cause Remedy
Electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Moisture in the plug Dry plug
Not enough electrolyte Top up electrolyte
Electrode obsolete Replace electrode
Electrode broken Replace electrode
Socket damp Dry socket
Calibration procedure:
Incorrect solution temperature
(without temperature sensor)
Set up correct temperature
Incorrect buffer solutions Select buffer solutions
suitable for the calibration
procedure
Buffer solutions too old Use only once.
Note the shelf life
Cause Remedy
pH electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Test sample
pH value not stable Measure with air excluded if
necessary
Temperature not stable Temper if necessary
Electrode + test sample:
Conductivity too low
(e.g. in ultrapure water)
Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode