User Manual

What to do if... pHotoFlex / pHotoFlex Turb
90
ba75491d04 07/2006
Obviously incorrect
measured values
Sensor symbol flashes
6.4 Turbidity
Error message
Measured values
obviously incorrect
Measured value display
< 0.01 FNU
Cause Remedy
pH electrode:
Not connected Connect electrode
Cable broken Replace cable or electrode
pH electrode unsuitable Use suitable electrode
Temperature difference
between buffer and test
sample too high
Adjust temperature of buffer
or sample solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Cell not correctly inserted Lock cell into place
Cell contaminated Clean the cell
Calibration too old Carry out calibration
Cause Remedy
Measured value outside the
measuring range
not possible