General AQA2
settings
Note
When the mode (Weeks or Measurements) is changed, all AQA2 intervals
are reset to the preset values.
Locking the method Here you configure whether or not a method will be locked against
measurement if there is no valid AQA2 check or the interval for the AQA2
check has expired.
1 Select and confirm Mode.
The Mode selection field pops up.
2 Select and confirm Weeks or
Measurements.
AQA2 is active. For all methods,
the AQA2 intervals are entered
either in weeks or number of
measurements.
3 Accept the general settings with
[Apply].
<HOME>
[AQA]
AQA2 setup
1 Select and confirm Lock methods.
2 Select and confirm Yes.
The method lock is enabled.
Each method will be locked if the
AQA2 check is invalid or the AQA2
interval has expired.
<HOME>
[AQA]
AQA2 setup