Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

photoLab
®
6100 VIS Operation
127
ba75847e01 08/2009
General AQA2
settings
Note
When the mode (Weeks or Measurements) is changed, all AQA2 intervals
are reset to the preset values.
Locking the method Here you configure whether or not a method will be locked against
measurement if there is no valid AQA2 check or the interval for the AQA2
check has expired.
1 Select and confirm Mode.
The Mode selection field pops up.
2 Select and confirm Weeks or
Measurements.
AQA2 is active. For all methods,
the AQA2 intervals are entered
either in weeks or number of
measurements.
3 Accept the general settings with
[Apply].
<HOME>
[AQA]
– AQA2 setup
AQA2 setup
04/16/07 9:52
Mode Weeks
Lock methods Yes
Method...
Method list
1 Select and confirm Lock methods.
2 Select and confirm Yes.
The method lock is enabled.
Each method will be locked if the
AQA2 check is invalid or the AQA2
interval has expired.
<HOME>
[AQA]
– AQA2 setup
AQA2 setup
04/16/07 9:52
Mode Weeks
Lock methods Yes
Method...
Method list