Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

Operation photoLab
®
6100 VIS
128
ba75847e01 08/2009
Activating AQA2
monitoring for a
method
Defining the AQA2
Interval, nominal
value and tolerance
The AQA2 Interval defines the interval between two AQA2 checks.
When an interval has expired, the following consequences become effective:
z Warning and loss of the AQA2 labeling
z Locking of the method against measurement (if activated).
Setting range:
1 to 12 weeks (default: 12 weeks) or
1 to 10000 measurements (default: 200 measurements)
Note
The unit of the AQA2 interval (Weeks or Measurements) is defined in the line,
Mode (see page 127).
Repeat the steps 1 to 8 if you want to configure further tests for AQA2.
1 Select a method (see section
4.5.3).
2 Select and confirm AQA2.
3 Select and confirm AQA2 active.
AQA2 is active for this method.
<HOME>
[AQA]
– AQA2 setup
– Method...
AQA2 setup
04/16/07 9:52
Method
3: A6/25
AQA2 AQA2 active
Interval 12 Weeks
Target value 4.00 mg/l NH
4
-N
Tol eranc e 0. 5 0 m g / l NH
4
-N
Standard ID
Method list
4 Select the Interval and enter the
AQA2 interval.
5 If necessary, adjust the values for
Target value and Tolerance.
6 Optional: Select Standard ID and
enter a designation. The
designation is recorded in the
AQA2 documentation.
AQA2 setup
04/16/07 9:52
Method
3: A6/25
AQA2 AQA2 active
Interval 12 Weeks
Target value 4.00 mg/l NH
4
-N
Tol eranc e 0. 5 0 m g / l NH
4
-N
Standard ID
Method list