Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

photoLab
®
6100 VIS Operation
133
ba75847e01 08/2009
7 In theStandard ID entry field,
select the simplified MatrixCheck
with the CombiCheck standard
solution or enter a designation for
another standard solution used.
If the CombiCheck is selected, no
more entries are required
(continue with step 10).
8 Enter the concentration of the
used standard solution in the
Standard concentration entry field.
Specifying the series of
measurements:
9 Enter the volumes of sample and
standard of the individual test
sample solutions in the columns,
Sample [ml] and Standard [ml].
The nominal value is calculated
after each entry.
- You can delete a measurement
from the series of
measurements with [Delete].
Note that all nominal values have
to be within the measuring range
of the test.
10 Using [Next], accept all entries on
the page and switch to the next
page. The entries are checked by
the photometer.
The photometer is ready to carry
out the series of measurements.
Carrying out the series of
measurements:
According to the program, the
samples are measured top down.
You can, however, select the
samples yourself and thus change
the order with <▲><▼>.
11 Use [Measurement] to proceed to
the measurement of the (first)
sample.
MatrixCheck (Spike)
04/16/07 9:52
Method 1: C3/25
Sample concentration 45 mg/lCOD
Standard ID COD 1500
Standard concentration 400 mg/lCOD
Sample Standard Target value
[ml] [ml] [mg/l]
10 0.5 62
10 1 77
10 1.5 91
Dilute Delete Next
MatrixCheck (Spike)
04/16/07 9:52
Method 1: C3/25
Sample concentration 45 mg/lCOD
Sample Standard Target value nominal
[ml] [ml] [mg/l] [mg/l]
10 0.5 62 58
10 1 77
10 1.5 91
Back Measureme Complete