Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

Operation photoLab
®
6100 VIS
134
ba75847e01 08/2009
The measurement display
appears.
12 Insert the cell with the respective
sample.
The sample is measured.
After the measurement, the
recovery rate is displayed in the
right table column.
The assessment of the recovery
rate is displayed next to
the recovery rate
(✓ or ✗
).
The criteria for the assessment
are determined in the menu,
AQA3/MatrixCheck setup /
Maximum difference.
13 If necessary, repeat the steps 11
and 12 for the remaining samples.
14 Use [Complete] to complete the
MatrixCheck.
The Save dialog box pops up.
15 If necessary, change the storage
location with [Location]:
Internal DataB folder:
Exchange folder in the instrument
or
USB memory:
USB memory device connected at
the USB-A connection.
16 If necessary, change the file
name.
17 Save the file with
<START·ENTER>.
MatrixCheck
04/16/07 9:52
Method 1: C3/25
Sample concentration 45 mg/lCOD
Sample 10 ml
Standard 0.5 ml
To start measurement,
insert cell or press <Start/Enter>
16 mm
Back
MatrixCheck
04/16/07 9:52
Method 1: C3/25
Sample concentration 45 mg/lCOD
Sample Standard Target value nominal
[ml] [ml] [mg/l] [mg/l]
10 0.5 62 58 94 %
10 1 77
10 1.5 91
Back
Measurement
Complete