Test record The result of the MatrixCheck is displayed in a test record. You can print this
record and save it as a file.
To save the file in the photometer, select the Internal DataB folder as the
location. To save the file in an external USB memory device at the USB-A
connection, select USB memory as the location (see section 4.11.1).
Sample printout:
The display returns to the
measured value display of the
original sample without spiking /
dilution.
The [MC] status indicator is
displayed. A MatrixCheck was
carried out for this measured
value.
SpectroFlex 6100 09130512 1.30-WTW-1.60 Administrator
MatrixCheck OK
Protocol ID 7
Method 1: C3/25 COD
Sample concentration 45 mg/lCOD
Standard ID COD 1500
Standard concentration400 mg/lCOD
Sample Standard Target value Actual value
ml ml mg/l mg/l
10 0.5 62 58 94% OK
10 1 77 71 92% OK
10 1.5 91 77 85% OK