Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

photoLab
®
6100 VIS Maintenance and cleaning
151
ba75847e01 08/2009
5 Maintenance and cleaning
5.1 Replacing the lamp
Service life of the
tungsten halogen
lamp
The tungsten halogen lamp is a wear part with a certain average service life
(see chapter 7 T
ECHNICAL DATA). It has to be replaced if defective. The
photometer has a service hour counter for the lamp module (see section
4.18).
Note
The replacement lamp is readily assembled as a lamp module and optically
adjusted in the factory. Therefore, treat it with utmost care. Fingerprints on
the glass will shorten the service life of the lamp. Do not touch the bulb of the
new lamp module with bare fingers. If you have touched the bulb
inadvertently, carefully clean it with a clean cloth soaked in alcohol.
Replacing the lamp
module
The lamp module is behind a cover (aluminium sheet) on the rear panel of
the photometer. Proceed as follows:
Caution
The lamp becomes very hot during operation. Do not touch the hot lamp
because it can cause burns! The lamp should cool down for approx.
10 minutes before it is exchanged.
1 Switch off the photometer and
disconnect it from the power
supply.
2 Unscrew the two screws (1) and
remove the lamp cover (2).
1
2