Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

photoLab
®
6100 VIS What to do if ...
157
ba75847e01 08/2009
Obviously incorrect
measured values
Fluctuating
measured values
Self test failed.
Connected printer
does not print
Cause Remedy
– Cell contaminated – Clean the cell
– Dilution set incorrectly – Set the dilution
– Selected method not suitable – Select different method
– Zero measurement incorrect – Perform zero measurement
– Blank value incorrect – Remeasure the blank value
– Cells that are not recognized (e.g.
some plastic cells) disturb the
AutoCheck measurement and, as
a consequence, falsify measured
values until the next AutoCheck is
carried out
– Use suitable cells (see section
7.1 and section 8.1)
Cause Remedy
– Cell shaft cover open – Close the cell shaft cover
Cause Remedy
– System test: Instrument defective – Return instrument to service
department
– Filter test: Instrument defective – Return instrument to service
department
– Lamp test: Lamp defective – Replace lamp (see section 5.1)
– Return instrument to service
department
– Wavelength calibration:
– Foreign particle in the cell shaft
– Instrument defective
– Clean the cell shaft
– Return instrument to service
department
Cause Remedy
– Printer not suitable – Connect a printer that can
interpret the printer control
language PCL-3