Owner's manual

Table Of Contents
photoLab
®
6100 VIS What to do if ...
157
ba75847e01 08/2009
Obviously incorrect
measured values
Fluctuating
measured values
Self test failed.
Connected printer
does not print
Cause Remedy
Cell contaminated Clean the cell
Dilution set incorrectly Set the dilution
Selected method not suitable Select different method
Zero measurement incorrect Perform zero measurement
Blank value incorrect Remeasure the blank value
Cells that are not recognized (e.g.
some plastic cells) disturb the
AutoCheck measurement and, as
a consequence, falsify measured
values until the next AutoCheck is
carried out
Use suitable cells (see section
7.1 and section 8.1)
Cause Remedy
Cell shaft cover open Close the cell shaft cover
Cause Remedy
System test: Instrument defective Return instrument to service
department
Filter test: Instrument defective Return instrument to service
department
Lamp test: Lamp defective Replace lamp (see section 5.1)
Return instrument to service
department
Wavelength calibration:
Foreign particle in the cell shaft
Instrument defective
Clean the cell shaft
Return instrument to service
department
Cause Remedy
Printer not suitable Connect a printer that can
interpret the printer control
language PCL-3