Owner's manual

Table Of Contents
photoLab
®
6100 VIS Operation
53
ba75847e01 08/2009
The photometer is ready to
measure.
1 Open the setting menu with
[Setup].
2 Select and confirm Sample blank
value.
3 Insert the cell with a suitable blank
sample.
The first single measurement for
the sample blank value takes
place.
The following data is displayed as
the result:
The measured absorbance
from the (last) single
measurement.
The median from all single
measurements carried out up to
now.
4 If necessary, carry out further
single measurements for the
formation of the median with [Next
meas.]
or
discard the last single
measurement with [Discard].
5 To accept the median value, press
[Apply].
The photometer is ready to
measure.
The use of the sample blank value
is indicated by [SB] in the top right
corner of the display.