Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

photoLab
®
6100 VIS Operation
71
ba75847e01 08/2009
4.6.3 Measuring against the Reference absorbance
Each time the photometer is switched on, the absorbance or transmission is
measured against the absorbance of the zero adjustment as a basis. You
can, however, also determine a Reference absorbance and use it as the
basis.
The Reference absorbance refers to the adjusted wavelength. The measured
value remains stored until
z the photometer is switched off
z the cell type is changed
z the wavelength is changed
z a new reference value is measured
z it is deleted manually ([Reference] / Delete).
z the Absorbance / % Transmission measuring mode is exited
Single and multiple
determination
The Reference absorbance can be determined with single or multiple
determination. With multiple determination, the mean value is calculated as
the median from the individual measured values.
Measuring the
Reference
absorbance
The settings of the last
measurement are active.
1 Start the reference measurement
with [Reference].
If a value for the reference
absorbance is already stored, it
can be deleted or overwritten by a
new reference measurement.
After the reference absorbance
value has been deleted, the
photometer measures against the
absorbance of the zero
adjustment.
<HOME>
Absorbance / % Transmission
Absorbance
04/16/07 9:52
To start measurement,
insert cell or press <Start/Enter>
489 nm
10 mm
Setup Wavelength Transmission Reference