Owner's manual
Table Of Contents
- 1 Overview
- 2 Safety instructions
- 3 Commissioning
- 4 Operation
- 4.1 Switching on or off the photometer
- 4.2 General operating principles
- 4.3 Photometer settings and system administration
- 4.4 Zero adjustment
- 4.5 Measuring in Concentration mode
- 4.5.1 Measuring cell tests with barcode
- 4.5.2 Measuring reagent tests with AutoSelector
- 4.5.3 Measuring reagent-free tests and user-defined methods
- 4.5.4 Exceeding the upper or lower limits of the measuring range
- 4.5.5 Selecting a method manually
- 4.5.6 Settings for Concentration mode
- 4.5.7 Measuring diluted samples
- 4.5.8 Sample blank value
- 4.5.9 Reagent blank value
- 4.5.10 Automatic Turbidity correction
- 4.5.11 Programming / modifying user-defined methods
- 4.5.12 The IQ LabLink procedure
- 4.6 Measuring the Absorbance / % Transmission
- 4.7 Multi wavelengths methods
- 4.8 Spectrum
- 4.9 Kinetics
- 4.10 Timer
- 4.11 Memory
- 4.11.1 Overview
- 4.11.2 Instructions on using USB memory devices
- 4.11.3 Measurement datasets
- 4.11.4 Saving measurement datasets manually
- 4.11.5 Saving measurement datasets automatically
- 4.11.6 Displaying measurement data memory
- 4.11.7 Filtering measurement datasets
- 4.11.8 Inverting filters
- 4.11.9 Erasing stored measurement datasets
- 4.12 Copying files
- 4.13 Transmitting data
- 4.14 Analytical quality assurance (AQA)
- 4.15 User management
- 4.16 Reset
- 4.17 Photometer information ([Info])
- 4.18 Lamp counter
- 4.19 Software and methods update
- 5 Maintenance and cleaning
- 6 What to do if ...
- 7 Technical data
- 8 Accessories and options
- Appendix

photoLab
®
6100 VIS Operation
83
ba75847e01 08/2009
Rectangular cell:
Open the inner turn-up lid.
Insert the rectangular cell
vertically so it touches the bottom
and left edge of the cell shaft. The
opaque sides of the rectangular
cell must point to the front and
back.
9 Close the cell shaft cover.
10 Start the measurement with
<START·ENTER>.
11 Wait until the spectrum is
completely recorded.
At the end of the recording the
following message appears:
Recording of spectrum is
completed.
12 Confirm the message with
<START·ENTER>.
The cursor appears at the
absolute maximum of the
spectrum.
13 You have the following options:
- Immediately edit the spectrum
(see section 4.8.3)
-With <PRINT>, you can output
the spectrum to a connected
printer as a graphic.
- You can save the spectrum as a
*.csv file with <STORE>. As the
storage location, you can select
the photometer (Internal DataB
folder) or a USB memory device
connected to the USB-A
connection (USB memory).
Stored spectra can be recalled
and edited at any time (see
section 4.8.3).
Spectrum
04/16/07 9:52
Wavelength [nm]
Setup Edit Zoom Open
Absorbance
Spectrum
04/16/07 9:52
Wavelength [nm]
Setup Edit Zoom Open
Absorbance
Cursor information