Owner's manual

Table Of Contents
Operation photoLab
®
6100 VIS
92
ba75847e01 08/2009
4.9.3 Recording the Kinetics
Note
During the recording, the photometer cannot carry out any regular self-test or
self-calibration (AutoCheck), because the recording would have to be
interrupted for this. A warm-up time of at least two hours is required for the
photometer to measure reliably during the recording.
Note the warm-up time of at least
2 hours for kinetic recordings.
1 Select the required profile with
[Profile] (see section 4.9.2).
The photometer is ready to
measure after the profile has been
selected.
2 Depending on the type, insert the
cell as follows:
Round cell:
Insert the round cell in the round
cell shaft so it touches the bottom.
If the inner turn-up lid is opened
too wide, a message prompts you
to close the inner turn-up lid.
<HOME>
Kinetics
inner
turn-up lid