Manual
Operation ProLab 1000
22
ba75576e04 01/2011
z Functions
Functions are indicated by the name of the function. They are
immediately carried out by confirming with <MENU/OK>.
Example: Display the Calibration record function.
Messages Information or operating instructions are designated by the [i] symbol.
They cannot be selected.
Example:
Note
The principles of navigation are explained in the two following sections
by reference of examples:
z Setting the language (see page 23)
z Setting the date and time (see page 25)
Calibration record
Buffer
: NIST/DIN
Calibration interval: 7 d
Unit for zero point pH
Unit for slope
: mV/pH
[i] 2.00 4.01 7.00 10.01
pH
[i] Buffer series NIST/DIN
[i] Immerse sensor in buffer 1
Set temperature: 25 °C
Continue
pH