Manual

Operation ProLab 1000
74
ba75576e04 01/2011
4.8.4 Operation with MultiLab pilot
With the aid of the MultiLab pilot software, you can record and evaluate
measuring data with a PC. The data is transmitted after the meter was
connected to the RS232 serial interface (COM port) or the USB
interface of the PC.
Note
More detailed information can be found in the MultiLab pilot operating
manual.
4.9 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.9.1 Resetting the sensor settings
Note
The calibration data is reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH The following settings for pH measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
respective measuring menu. To open it, activate the relevant
measuring window in the measured value display and press the
<MENU/OK> key.
Setting Default settings
Cal. interval 7 d
Unit for slope mV/pH
Measured parameter pH
High resolution pH On
High resolution mV On
Zero point 0 mV
Slope -59.16 mV
Temperature, manual 25 °C