User guide

ProLab 1000 Operation
49
ba75576e05 11/2013
Continuing with three-
point to five-point
calibration
(Buffer NIST/DIN)
12 Wait for the measurement with stability control to be completed
or Terminate stability control with <MENU/OK> and accept the
calibration value.
The calibration display for the next buffer appears.
13 For two-point calibration, select Exit with 2 point calibration with
<><> and confirm with <MENU/OK>.
The calibration is completed as a two-point calibration.
The new calibration values are displayed as a message ([i]).
You have the following options:
Accept the new calibration values with <MENU/OK>.
Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.
To switch to the measured value display without
accepting
the new calibration values, press <MODE> or <ESC>.
14 Thoroughly rinse the electrode with distilled water.
15 Immerse the electrode in buffer solution 3.
16 If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <><> and <MENU/OK> in the
Set temperature setting.
17 Select Continue with <><> and press <MENU/OK>. The
buffer is measured.
The measured value is checked for stability (Stability control).