Manual

What to do if... ProLab 2000
100
ba75577e05 11/2013
No stable measured
value
Obviously incorrect
measured values
Buffer solutions:
Incorrect buffer solutions
Change buffer solutions
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change buffer solutions
Cause Remedy
Cause Remedy
pH electrode:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low Use a suitable electrode
Temperature too high Use a suitable electrode
Organic liquids Use a suitable electrode
Cause Remedy
pH electrode:
Not connected Connect electrode
Cable broken Replace electrode
pH electrode unsuitable Use a suitable electrode
Temperature difference
between buffer and test
sample too high
Adjust temperature of buffer
or sample solutions