ProLab 2000 What to do if...
101
ba75577e05 11/2013
CalClock flashes
6.2 ISE measurement
Error message Err1
Obviously incorrect
measured values
Error message, Error
(invalid
calibration)
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Measuring range exceeded Dilute test sample
Cause Remedy
Electrode not connected Connect electrode
Cable broken Replace electrode
Cause Remedy
ISE electrode:
Moisture in the plug Dry plug
Electrode obsolete Replace electrode
Electrode unsuitable for the
range to be measured
Use suitable electrode
Socket damp Dry socket
Calibration procedure:
Wrong sequence of standards
for three point calibration
Select correct sequence
Calibration standards do not
have the correct temperature
(max.
± 2 °C temperature
difference)
Adjust the temperature of
the calibration standards