ProLab 2000 What to do if...
103
ba75577e05 11/2013
Error message, OFL
Error message, Error
(invalid
calibration)
Error message, LEAK
Measured value flashes
for a long time
(No stable measured
value)
Measured value too low
Measured value too
high
Cause Remedy
Measuring range exceeded not possible
Cause Remedy
D.O. sensor:
Electrolyte solution depleted Regenerate probe
Membrane contaminated Clean membrane
Electrode system poisoned Regenerate probe
Cause Remedy
D.O. sensor:
Membrane damaged Exchange the membrane cap
Membrane head screwed on
not tight enough
Screw membrane head tight
Cause Remedy
D.O. sensor:
Membrane contaminated Clean membrane
Cause Remedy
D.O. sensor:
Insufficient flow Provide flow to the probe
Cause Remedy
High amount of dissolved
substances
Correct solubility function
using the salinity equivalent
Air bubbles bump on the
membrane with high velocity
Avoid direct flow to the
membrane