Manual

ProLab 2000 Operation
47
ba75577e05 11/2013
Sample record:
Note
If non-ID sensors are used, the sensor type and series number are not
output in the calibration record.
Calibration evaluation After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.
Display Calibration
record
Zero point [mV] Slope
[mV/pH]
+++ -18 ... +18 -60.5 ... -57.5
++ -22 ... +22 -57.5 ... -56
+ -26 ... +26 -61 ... -60.5
or
-56 ... -55
- -30 ... +30 -62 ... -61
or
-55 ... -50
Clean the electrode according
to the sensor operating manual
CalError CalError
< -30 or
> 30
... -62 or
... -50
Eliminate the error according to
chapter 6 W
HAT TO DO IF...
(page 99)