Manual
Operation ProLab 2000
88
ba75577e05 11/2013
Sample printout
Menu item Setting/
function
Description
Data storage /
Calibration data
storage /Display
- Displays the calibration records.
Further options:
Scroll through the calibration
records with <▲><▼>.
Output the displayed
calibration record to the
interface with <PRINT>.
Quit the display with <ESC>
or <MENU/OK>.
Switch directly to the
measured value display with
<MODE>.
Data storage /
Calibration data
storage / Output to
RS232/USB
- Outputs to the interface the
calibration records.
Printing date 26.04.06 16:13
ProLab 2000
Ser. no. 06249876
USER ID: 1234567890
CALIBRATION pH
Calibration date 20.04.06 10:14:03
A 161 1M DIN-ID
Ser. no. A062498765
USER ID: 1234567890
Cal. interval 7 d
DIN
Buffer 1 6.86
Buffer 2 9.18
Voltage 1 7.2 mV 26.3 °C
Voltage 2 -171 mV 26.3 °C
Slope -59.2 mV/pH
Zero point -0.5 mV
Sensor +++
________________________________________
CALIBRATION ISE
Calibration date 20.04.06 10:14:03
Standard 1 0.010 mg/l
Standard 2 0.020 mg/l
Voltage 1 38.5 mV 25.0 °C
Voltage 2 58.0 mV 23.0 °C