User guide
ProLab 3000 System functions
65
ba75645e02 01/2010
6.10 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
6.10.1 Resetting the sensor settings
The sensor settings are reset in the sensor menu.
In the measured value display, activate the relevant sensor window
[Tab] and open the menu with [Menu].
pH The following settings for pH measurements are reset to the default
settings with the Reset pHMeasuring function:
6.10.2 Resetting the system settings
The following system settings are reset to the default settings with the
System / Reset function:
Note
The calibration data is reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
Setting Default settings
Calibration interval... 7 d
Slope in mV/pH
Measured parameter pH
pH resolution High
Zero point 0 mV
Slope -59.16 mV/pH
Temperature, manual 25 °C
Setting Default settings
Language English
Temperature unit Celsius
Signal On
✓
Baud rate (Baud) 4800
Output format ASCII