Manual

Conductivity ProLab 4000
150
ba75646d02 01/2010
Sample record:
Calibration evaluation After the calibration, the meter automatically evaluates the current sta-
tus of the calibration. The evaluation appears on the display and in the
calibration record.
Print date 26.04.07 16:13
ProLab 4000 (06249876)
Current user: 1234567890
CALIBRATION Cond
Calibration date 20.04.07 10:14:03
LF413-ID (A062498765)
User: Administrator
Calibration interval: 150 days
Standard: 0,01 mol/l KCl
Cond./Tref 25: 1413 µS/cm
Cal. temp: 25,4 °C
Cell constant: 0.475 1/cm 25.0 °C
Sensor +++
Note
If non-ID sensors are used, the sensor type and series number are not
output in the calibration record.
Display Calibration record Cell constant [cm
-1
]
+++ within the ranges
0.450 ... 0.500 cm
-1
0.585 ... 0.715 cm
-1
or
0.800 ... 1.200 cm
-1
Error Error
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
Eliminate the error according to chapter
16 W
HAT TO DO IF... (page 189)