Overview ProLab 4000
18
ba75646e02 01/2010
1.5.2 Sensor data from ID sensors
ID sensors transmit the following sensor data to the meter:
z SENSOR ID
Sensor type
Sensor series number
z Calibration data
Calibration date
User (having carried out the last calibration procedure)
Calibration characteristics
Calibration interval
Selected buffer set (pH electrodes only)
Last cell constant (conductivity measuring cell only)
z Measurement settings (conductivity measuring cells only)
Specified reference temperature
Adjusted temperature coefficient
Adjusted TDS factor
The calibration data is updated in the ID sensor after each calibration
procedure. A message is displayed while data is updated in the sensor.
Note
While data is updated in the sensor, the sensor must not be discon-
nected because otherwise the calibration data will not be completely
transmitted. The sensor will then have no valid calibration.
Note
If non-ID sensors are used, the calibration data from the meter is used
and also stored in the meter.