Warning [TempErr]
Warning [ISEErr]
16.3 Conductivity measurement
Conductivity measuring
cell was not recognized
Error message, Error
(invalid calibration)
CalClock
with red frame
Cond temp?
Cause Remedy
Temperature difference
between measurement and cali-
bration greater than 2 K.
Adjust the temperature of
the test sample
Cause Remedy
Electrode voltage outside cali-
brated range
Recalibrate
Cause Remedy
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and replace it if
necessary
Unsuitable control standard Use control standard,
0.01 mol/l KCl
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Test sample does not have the
correct temperature (max.
± 2
°C temperature difference from
calibration temperature)
Adjust the temperature of the
test sample