Manual

Overview ProLab 4000
18
ba75646e03 12/2012
1.5.2 Sensor data from ID sensors
ID sensors transmit the following sensor data to the meter:
SENSOR ID
Sensor type
Sensor series number
Calibration data
Calibration date
User (having carried out the last calibration procedure)
Calibration characteristics
Calibration interval
Selected buffer set (pH electrodes only)
Last cell constant (conductivity measuring cell only)
Measurement settings (conductivity measuring cells only)
Specified reference temperature
Adjusted temperature coefficient
Adjusted TDS factor
The calibration data is updated in the ID sensor after each calibration
procedure. A message is displayed while data is updated in the sensor.
Note
While data is updated in the sensor, the sensor must not be discon-
nected because otherwise the calibration data will not be completely
transmitted. The sensor will then have no valid calibration.
Note
If non-ID sensors are used, the calibration data from the meter is used
and also stored in the meter.